Справочник Пользователя для Boston Scientific Corporation CRM293604
14
Voice Messages
Confirm request to clear all records
Action: You pressed the blue PRINT button
and green INTERROGATE
button
; the HF PARTNER device will ask you to confirm that you want to
delete all records. Press the blue PRINT button
within 10 seconds of this
voice message to confirm that you want to delete all records.
Battery Low
Action: You should replace the HF PARTNER device batteries. This voice
message will be heard with the speaker control turned On or Off.
message will be heard with the speaker control turned On or Off.
Speaker Off
Action: You turned the speaker Off.
PARTNER Fault Detected
Action: The HF PARTNER device failed a self-test due to a reason other than
depleted batteries. A system status report can be printed from a qualified printer.
You should call Guidant Technical Services to report the problem. See
“Contacting Guidant” on page 5.
depleted batteries. A system status report can be printed from a qualified printer.
You should call Guidant Technical Services to report the problem. See
“Contacting Guidant” on page 5.
Language list
Action: You pressed the green INTERROGATE button
once while the
HF PARTNER device was in Configuration Mode. This accessed the list of
available languages. Now you can select your preferred language by pressing the
blue PRINT button
available languages. Now you can select your preferred language by pressing the
blue PRINT button
Printer paper error
Action: The printer is out of paper or there is a paper jam. If it is out of paper,
reload paper. If the paper is jammed, clear the paper path of any obstruction.
reload paper. If the paper is jammed, clear the paper path of any obstruction.
Printer toner error
Action: The printer is out of toner; replace the toner cartridge. This message will
also be heard when the wrong toner cartridge has been inserted.
also be heard when the wrong toner cartridge has been inserted.
Table 1. Voice Messages (Continued)