Справочник Пользователя для Boston Scientific Corporation CRM312004

Скачать
Страница из 38
11
Keys
For specific instructions on how to operate the PRM keys and use the te-
lemetry wand, refer to the System Guide for the Guidant pulse generator 
being interrogated. General key functions are summarized below.The key 
pad on the right side of the PRM is shown in Figure 6.
Table 1  Key Functions
Left side of key pad:
0 (stop), 10, 25, 50, 
and 100 mm/sec 
(paper-speed keys)
Press the speed keys to specify the paper speed for 
the internal printer/recorder. The printout will show 
the date and time, lead(s) being printed, gain setting, 
chart speed, and filter setting. To stop the printer/
recorder, press the speed key labeled “0” (zero).
Press the paper-feed key to scroll the printer paper 
on the internal printer/recorder.
Press this key to cause the internal printer/recorder 
to print a one-mV calibration pulse.
Press this key to force the trace back to the baseline 
after a defibrillation shock.
Right side of key pad
Press STAT PACE when in telemetry communication 
with the pulse generator to program predetermined 
parameters for emergency bradycardia pacing
Press DIVERT THERAPY when in telemetry 
communication with the pulse generator to divert 
tachycardia therapy delivery.
STAT PACE
DIVERT THERAPY