TROUBLESHOOTING
If the ZWT does not operate properly, check that the USB cable is securely connected to both the ZWT and the PRM. Also check that the PRM cords and 
cables are free of visible defects and are properly connected. Possible causes and corrective actions for ZWT problems are shown below.
Symptom
Possible Cause
Corrective Action
Green indicator light on ZWT 
does not light within 60 seconds 
of powering on PRM
USB cable not securely connected to ZWT 
and PRM
Remove and reconnect both ends of USB cable.  Reposition 
wand over the pulse generator and repeat interrogation.
USB cable damaged
Replace with Model 3141 USB Cable only.
ZWT fault
Contact Boston Scientifi c representative.
Intermittent or no RF telemetry 
communication
Telemetry RF signal obstructed
Assure that a clear line-of-sight path exists between ZWT 
and pulse generator. Repeat interrogation.
Telemetry RF signal interference
Reposition or reorient ZWT at least 7.6 cm (3 in)  away from 
PRM. Repeat interrogation.
USB cable not securely connected to ZWT 
and PRM
Remove and reconnect both ends of USB cable.  Reposition 
wand over the pulse generator and repeat interrogation.
RF Telemetry fails
Reposition wand over pulse generator; repeat interrogation.
PRM software version not current
Contact Boston Scientifi c representative.