Справочник Пользователя для Biolase Inc. EPIC-1
EPIC™10 User Manual
pg. 15
5400525 Rev B
Two (2) AAA batteries are required to power the footswitch (included). (For instructions on how
to replace the footswitch batteries, see Section 4.)
to replace the footswitch batteries, see Section 4.)
Figure 3.3: Footswitch
Remote Interlock
This feature allows the laser console to be connected to the remote sensor, preventing its
operation when triggered (e.g., by opening door). The electric cable from this connector should
be wired to the normally closed switch, sensing the opening of a door and turning the laser
console off when the switch is open.
operation when triggered (e.g., by opening door). The electric cable from this connector should
be wired to the normally closed switch, sensing the opening of a door and turning the laser
console off when the switch is open.
To override this feature, don’t connect the plug.
Figure 3.4: Remote Interlock Connector
Emergency Stop
Press the red Emergency Laser Stop button to instantly turn off the laser console. The error
screen will display an “Emergency Switch Error” message and the amber LED will begin
flashing. To clear the error, press the Emergency Laser Stop button again; in 2 to 5 seconds the
amber LED will stop flashing and the system will automatically go into STANDBY mode.
screen will display an “Emergency Switch Error” message and the amber LED will begin
flashing. To clear the error, press the Emergency Laser Stop button again; in 2 to 5 seconds the
amber LED will stop flashing and the system will automatically go into STANDBY mode.