EPIC™10 User Manual 
pg. 15 
5400525 Rev B  
 
 
Two (2) AAA batteries are required to power the footswitch (included).  (For instructions on how 
to replace the footswitch batteries, see Section 4.) 
 
 
 
Figure 3.3: Footswitch 
 
Remote Interlock 
This feature allows the laser console to be connected to the remote sensor, preventing its 
operation when triggered (e.g., by opening door). The electric cable from this connector should 
be wired to the normally closed switch, sensing the opening of a door and turning the laser 
console off when the switch is open. 
To override this feature, don’t connect the plug. 
 
Figure 3.4: Remote Interlock Connector 
 
Emergency Stop 
Press the red Emergency Laser Stop button to instantly turn off the laser console. The error 
screen will display an “Emergency Switch Error” message and the amber LED will begin 
flashing. To clear the error, press the Emergency Laser Stop button again; in 2 to 5 seconds the 
amber LED will stop flashing and the system will automatically go into STANDBY mode.