EPIC™10 User Manual 
pg. 32 
5400525 Rev B  
 
 
 
LASER 
WARNING: 
Never point the laser at a person’s eyes. All persons present in the 
operatory must wear protective eyewear when the laser is in 
operation 
 
Pre-programmed Settings for Dental Procedures 
To access the pre-programmed procedure values: 
1.  Go to the Procedures menu by pressing the Soft Tissue icon 
 on the Home screen. 
2.  Press the button associated with the desired procedure. 
3.  Press the up and down arrows 
 to scroll for additional procedures. 
CAUTION: 
Always use clinical judgment when selecting power, pulse, length, and pulse 
interval parameters to ensure optimal clinical results. The recommended settings 
apply only to the 300μm and 400μm tips. 
 
To store your personal preferred settings for any procedure: 
A.  Follow steps 1 and 2 above. 
B.  Enter the new values. 
C.  Touch and hold the Procedure name for more than 2 seconds; you will hear a beeping 
sound confirming the settings are saved. 
 
NOTE: 
The Procedure Pre-Sets installed at the factory are based on clinical 
recommendations and feedback from experienced laser dentists. 
 
300μm tips are recommended for removing thin tissue layers. 400μm tips are 
recommended for removing fibrous tissue. 
 
Always use your clinical judgment when selecting power, pulse length, and pulse 
interval parameters to ensure optimal clinical results. At all times observe the clinical 
effects on the treatment area and adjust parameters accordingly.