EPIC™10 User Manual 
pg. 36 
5400525 Rev B  
 
 
Manual Cleaning of the Surgical Handpiece: 
Cleaning must be performed within a maximum of 1 hour after the procedure and always 
prior to sterilization.
 
 
1.  After use, carefully remove the tip from the handpiece and dispose of in a biohazard 
medical waste Sharps container. 
2.  Carefully remove the handpiece from the fiber optic cable (see Section 2). 
3.  Prepare any commercially available surgical instrument detergent/enzymatic cleaning 
solution with a pH of 7.0, such as Enzol
®
 or similar enzymatic presoak and cleaner, per the 
manufacturer’s instructions. (Follow the manufacturer’s instructions for disposal of used 
solution.) 
4.  Rinse the Handpiece under running lukewarm tap water (22 – 43°C) 
for a 
minimum of 10 seconds
 to remove gross soil. 
5.  Wrap the handpiece in a piece of gauze that has been soaked in the cleaning solution; 
leave it wrapped in the gauze for 
minimum of 10 minutes
6.  Unwrap the handpiece from the gauze and use a soft-bristled brush dipped  in the 
cleaning solution to gently scrub it for at least 
15 seconds
7.  Rinse the handpiece under running lukewarm tap  water  ( 2 2 - 4 3 ° C )   for a minimum of 
10 seconds
 and then dry with  a lint-free cloth. 
8.  Visually inspect the  handpiece for  any residual soil.  If necessary, repeat  steps 5 - 7 until 
all residual soil is removed. 
 
Steam Sterilization for Surgical Handpiece, Single Use Tips  
The steam sterilization process is intended to destroy infectious microorganisms and 
pathogens. 
•  Place the handpiece and fiber tips in separate single-wrap, self-seal autoclave pouches. 
•  Place on an autoclave tray; do not stack other instruments on top of the pouches. 
•  Place the tray inside the autoclave chamber and set the appropriate cycle as 
recommended in Figure 8.1.