6.1 Cleaning
The NGP1 Computing Platform is designed for being subjected to periodic cleaning and
disinfection, as typical for infection control, in a healthcare environment, Materials and
shapes as such that the unit can be cleaned with sprays and liquids.
Recommended cleaning solutions have to follow specific institutional recommendations
and are depending on the environmental situations in which the NGP1 Computing Platform
is used.
disinfection, as typical for infection control, in a healthcare environment, Materials and
shapes as such that the unit can be cleaned with sprays and liquids.
Recommended cleaning solutions have to follow specific institutional recommendations
and are depending on the environmental situations in which the NGP1 Computing Platform
is used.
Turn the system off and disconnect the AC adapter before cleaning.
6.1.1 Tested substances
Liquids in hospitals, expected to contact/affect equipment can be divided in the following
groups:
• Disinfection / cleaning products on alcohol basis;
• Disinfection / cleaning products on alkali basis;
• Disinfection / cleaning products on water basis;
• Disinfection / cleaning products on chlorine basis;
• polluting agents;
• physiological fluids.
The NGP1 Computing Platform has been tested against rubbing force, long term exposure
and contrast liquid tests, with the following agents:
Liquids or fluids
Isopropanol 100% - CAS 67-630
Haemosol 1% in 1 litre water – 00-238-1
Chlorine 250 ppm in 1 litre distilled water
Chlorehexidine 0,5% in 70% Ethanol – 00-125
Ethanol 70% - 70-CAS 64-17-5
Cidex- Phthalaldehyde solution – 00-237
Iodine 1% in 70% Ethanol – 06-318-1
Artificial sweat (5% KCL, pH=3) – 00-422-1
NaCI solution 0.9% – sodium chloride 00-236
Hexabrix 320
Coffee with milk and sugar
Alcohol 95%
Isopropanol 100% - CAS 67-630
Haemosol 1% in 1 litre water – 00-238-1
Chlorine 250 ppm in 1 litre distilled water
Chlorehexidine 0,5% in 70% Ethanol – 00-125
Ethanol 70% - 70-CAS 64-17-5
Cidex- Phthalaldehyde solution – 00-237
Iodine 1% in 70% Ethanol – 06-318-1
Artificial sweat (5% KCL, pH=3) – 00-422-1
NaCI solution 0.9% – sodium chloride 00-236
Hexabrix 320
Coffee with milk and sugar
Alcohol 95%
Table 7 - Table of tested substances
WARNING: Small stain /affection has been observed after long exposure
test due to Iodine, Haemosol and Hexabrix 320. Clean the product
test due to Iodine, Haemosol and Hexabrix 320. Clean the product
immediately if exposed to this substances to avoid halos and
discoloration.
discoloration.
6.1.2 To clean the screen
To removes dust and other particles, wipe the screen with a clean, soft, lint-free cloth.
For heavy duty cleaning, wipe the screen with a soft, clean cloth lightly sprayed with
prescribed cleaning or disinfectant material that should not contain wax or abrasives.
prescribed cleaning or disinfectant material that should not contain wax or abrasives.
Preferably apply the cleaner to the cloth, not the screen.