Справочник Пользователя для ActiGraph LLC 01
www.theactigraph.com
P a g e
| 8
49 E. Chase St. Pensacola, FL 32502 tel 850.332-7900 fax 850-332-7904
400
Sunrise/sunset
1,000
Overcast day
10,000-25,000
Full daylight
32,000-130,000
Direct sunlight
Table 1 – Lux Level Interpretation Estimates
Steps
Step counts are accumulated on a per-epoch basis and are based on accelerometer data collected on
the vertical axis. An algorithm present in the device firmware filters out the accelerometer’s baseline
noise level to help accurately accumulate the steps-per-epoch.
the vertical axis. An algorithm present in the device firmware filters out the accelerometer’s baseline
noise level to help accurately accumulate the steps-per-epoch.
Inclinometer
The inclinometer feature can help users identify the orientation of the device and, more importantly,
when the device itself was taken off. Each epoch is flagged with a number (1 through 4) to indicate the
orientation of the device during that epoch. For wGT3X+ and wActiSleep+ devices, this is done within
the ActiLife software when converting the *.gt3x file to *.agd format. As noted, the inclinometer
feature is only valid when the device is worn vertically on the point of the hip. More details are
available in the
when the device itself was taken off. Each epoch is flagged with a number (1 through 4) to indicate the
orientation of the device during that epoch. For wGT3X+ and wActiSleep+ devices, this is done within
the ActiLife software when converting the *.gt3x file to *.agd format. As noted, the inclinometer
feature is only valid when the device is worn vertically on the point of the hip. More details are
available in the
Important The inclinometer feature is only valid if the device is worn on the hip
with Axis 1 upward facing.
with Axis 1 upward facing.
Inclinometer Code
(Stored with each Epoch)
Interpretation
0
Device Off (Not Being Worn)
1
Subject Standing
2
Subject Lying Horizontal
3
Subject Sitting
Table 3 - Inclinometer Definitions
Off Indication on wGT3X+ and wActiSleep+ versus other devices
The Inclinometer feature offers researchers the ability to detect periods of time during which the device
was not worn. For devices other than the wGT3X+, wActiSleep+, GT3X+, and ActiSleep+, the
inclinometer algorithm makes this prediction by examining the angle at which the device is placed. In
the wGT3X+, wActiSleep+, GT3X+, and ActiSleep+ devices, off detection is performed by analyzing both
the angle at which the unit is placed as well as the length of time the device remains still. Absolute
was not worn. For devices other than the wGT3X+, wActiSleep+, GT3X+, and ActiSleep+, the
inclinometer algorithm makes this prediction by examining the angle at which the device is placed. In
the wGT3X+, wActiSleep+, GT3X+, and ActiSleep+ devices, off detection is performed by analyzing both
the angle at which the unit is placed as well as the length of time the device remains still. Absolute