8
[2] Turn the power switch to "ON".
When the power is turned ON, the SpO
2
measurement mode is in
continuous mode. To change the mode to intermittent mode, press the
mode switch. In the intermittent mode, the SpO
mode switch. In the intermittent mode, the SpO
2
will be measured at
30 seconds interval and n o pulse waveform is sent, but ECG and
respiration will be continuously measured. In the continuous SpO
respiration will be continuously measured. In the continuous SpO
2
measurement mode, the SpO
2
value and pulse waveform will be
continuously measured.
CAUTION
In the intermittent mode, the SpO
2
is measured at 30 seconds interval.
Even when the SpO
2
value suddenly changes during this interval, the
transmitter will only show the last SpO
2
measurement value. For the
patient with the possibility of sudden change, select the continuous
mode.
mode.
In the intermittent mode, the battery life is about 5 days and about 3
days in continuous mode.
days in continuous mode.