Справочник Пользователя для Fukuda Denshi Co Ltd LX5630

Скачать
Страница из 41
15 
nTG-26 (UD-5C) Attachment 
The TG-26 (UD-5C) is suitable for long-term measurement and can be 
attached to a finger of an adult, or to a foot, palm, wrist or ankle of a 
neonate.  Before attaching the probe to a patient, clean the skin using a 
cloth moistened with sterilizing alcohol. 
 
 
           
 
 
 
CAUTION 
Precautions for Use of TG-26 (UD-5C) 
· 
Cover the probe with a black cover if it is exposed to direct sunlight.    Some 
measurement error may occur if the probe is exposed to direct sunlight during 
measurements. 
· 
Make sure that the two pads are aligned . 
· 
If  the use of double-sided adhesive tape causes skin irritation, stop using it.  In 
such case, try using a bandage  tape. 
· 
Attach the pad with the LED to the base of  the fingernail and the sensor pad to 
the opposite side of the finger.  
· 
If measurements are taken over a long period of time, change the measuring site 
several times a day to prevent rash, redness or a low-temperature burn.  Be 
especially careful for the  continuous use  on premature infant, neonate, or a 
patient  with peripheral circulatory disturbance by changing the measuring site 
frequently.   
· 
When fixating the probe by bandage tape, do not tighten it too hard as this may 
cause edema or congestion.     
· 
If the  use of adhesive tape causes skin irritation, it should be discontinued.
 
· 
Since pediatric patients skin can be extremely sensitive, be careful when removing 
the tape.
 
· 
Inaccurate measurements can be caused by the placement of SpO
2
 probe on an 
extremity with a blood pressure cuff, arterial catheter,  or intravascular line.
 
 
LED 
Sensor 
TG-26 (UD-5C) 
TG-26 (UD-5C)