6. SpO
Monitoring (LX-7230N) 
18 
 
SpO
2
 Monitoring (LX-7230N) 
LX-7230N, which has a built-in Nellcor
®
 SpO
module, is described in this 
section. 
Regarding the LX-7230KM, which has a built-in Konica Minolta
®
 SpO
module, 
refer to page 35. 
 
< SpO
Sensors for LX-7230N > 
 
The optional SpO
sensors available for LX-7230N are as follows. 
The following table shows applicable patient and proper site for each SpO
sensor. Select the proper one depending on the purpose and intended use.   
 
Sensor Types 
Applicable Patient 
Applied Site 
OxiMax
® 
 MAX-I
 
Infant 
(weight of 3 to 20Kg) 
Toe 
OxiMax
®  
MAX-P 
Pediatric 
(weight of 10 to 50Kg) 
Finger 
OxiMax
®  
MAX-A/AL 
Adult 
(weight of 30Kg and over) 
Finger 
Durasensor
®  
DS-100A 
Adult 
(weight of 40Kg and over) 
Finger 
OxiMax
®  
MAX-R 
Adult  
(weight of 50Kg and over) 
Nose 
OxiMax
®  
MAX-FAST 
Adult/Pediatric 
(weight of 10Kg and over) 
Forehead 
Adult  
(weight of 40Kg and over) 
Finger 
OxiMax
®  
MAX-N 
Neonate  
(weight of less 3Kg) 
Foot 
 
WARNING 
 For SpO
2
 monitoring, always use the sensor specified by Fukuda 
Denshi. If any other sensor is used, high temperature rise of the sensor 
may place the patient in danger of burns in the worst case. 
  As with all medical equipment, carefully route cables to reduce the 
possibility of patient entanglement and strangulation.