6. SpO
2
 Monitoring (LX-7230N) 
21 
 
Reapplication 
 
1.  The MAX-I can be reused on the same patient as long as the adhesive 
tape attaches without slippage. 
2.  Enclosed adhesive “dots” are provided for reapplication. Place a 
transparent dot over each window as shown, and then remove the 
protective paper that covers each dot (Figure (5)). The sensor is now ready 
to be reapplied to the same patient. For the reapplication, do not remove 
the previous adhesive dot, but place the enclosed adhesive dot over it. 
 (5) 
 
CAUTION 
  Precautions for Use of adhesive sensor, MAX-I 
 Do not reuse the sensor on other patients. This is a sterilized product 
and it is intended for single patient use only. 
 Circulation distal on the sensor site should be checked routinely. The 
site must be inspected every 8 hours to ensure adhesion, skin 
integrity, and correct optical alignment. If skin integrity changes, move 
the sensor to another site 
 Do not use the sensor on patients who exhibit allergic reactions to the 
adhesive tape. 
 Failure to apply the sensor properly may cause incorrect 
measurements. 
 While the sensor is designed to reduce the effects of ambient light, 
excessive light may cause inaccurate measurements. In such cases, 
cover the sensor with an opaque material. 
 If the sensor is wrapped too tightly or supplemental tape is applied, 
venous pulsations may lead to inaccurate saturation measurements. 
 Excessive motion may compromise performance. In such cases, try to 
keep the patient still, or change the sensor site to one with less 
motion. 
 Intravascular dyes or externally applied coloring such as nail polish, 
dye, or pigmented cream may lead to inaccurate measurements. 
 Do not pull the sensor cable to remove the sensor from the device. 
 In the event of damage to the sterile packaging, do NOT use. Make 
sure to check whether the packaging and product is cracked or 
damaged before use. If there is any damage. 
 Do not immerse in water or cleaning solutions. Do not resterilize. 
 
For additional warnings, cautions or contraindications when using 
sensors with Nellcor-compatible device, refer to each SpO
2
 sensor 
instruction manual.