  Applying the OxiMax
®
 MAX-N sensor 
 
This Nellcor
®
 adhesive sensor, model MAX-N, is indicated for continuous 
noninvasive arterial oxygen saturation and pulse rate monitoring and can be 
reused on the same patient as long as the adhesive tape attaches without 
slippage. 
 
1.  Remove the plastic backing from the MAX-N and locate the two 
transparent windows on the adhesive side. Windows cover optical 
components. Note the corresponding alignment marks (a) on the non-
adhesive side and the dashed line (b) midway between the marks (Figure 
(1)). 
2.  Orient the MAX-N so that the dashed line is on the lateral edge of the site 
(a): 
 
Neonates: The preferred site is the foot. Alternatively, use the hand. The 
window next to the cable goes on the sole of the foot as shown (Figure (2)). 
 
Adults: 
The preferred site is the index finger. Alternatively, other fingers 
may be used. The window next to the cable goes on the nail side, distal to 
the first joint. Do not place on a joint. Note that the cable must be 
positioned on the top of the hand (Figure (3)). 
 
NOTE 
When selecting a sensor site, priority should be given to an extremity free of 
an arterial catheter, blood pressure cuff, or intravascular infusion line. 
 
3.  Wrap the MAX-N firmly, but not too tightly around the foot or finger. 
Windows must oppose each other. 
4.  Connect the MAX-N into the LX-7230N. Verify proper operation as 
described in the operation manual. 