Справочник Пользователя для Fukuda Denshi Co Ltd LX7230KM

Скачать
Страница из 112
6. SpO
2
 Monitoring (LX-7230KM) 
35 
 
SpO
2
 Monitoring (LX-7230KM) 
LX-7230KM, which has a built-in Konica Minolta
®
 SpO
module, is described in 
this section. 
Regarding the LX-7230N, which has a built-in Nellcor
®
 SpO
module, refer to 
page 18. 
 
< SpO
probes for LX-7230KM > 
 
The optional SpO
probes available for LX-7230KM are as follows. 
The following table shows applicable patient and proper site for each SpO
probe. Select the proper one depending on the purpose and intended use.   
 
Sensor Types 
Applicable Patient 
Applied Site 
Finger Clip 
SR-5C 
Adult and Pediatric 
(weight of 30Kg and over) 
Finger 
Spot Check 
SP-5C 
Adult and Pediatric 
(weight of 30Kg and over) 
Finger 
Universal Probe 
UD-5C 
Adult and Pediatric 
(weight of 40Kg and over) 
Finger 
Adult and Pediatric 
(weight of 15Kg and over) 
Finger 
Personal Probe 
SD-5C(25),  
SD-5C(20) 
Pediatric and Infant 
(weight of 5 to 15Kg) 
Big toe or the root 
of the big toe 
 
WARNING 
 For SpO
2
 monitoring, always use the probe/relay cable specified by 
Fukuda Denshi. If any other probe/relay cable is used, high temperature 
rise of the sensor may place the patient in danger of burns in the worst 
case. 
  As with all medical equipment, carefully route cables to reduce the 
possibility of patient entanglement and strangulation.