7. Measurement 
Turn ON the power and the measurement starts. 
 
 Starting Screen 
When the power is turned ON, the channel number configured on the LX-7230 
is displayed at the top of the LCD. 
Make sure whether the channel number on the 
LCD matches the channel number indicated on 
the label of the LX-7230 and the channel number 
configured on the receiving monitor. 
This screen automatically moves onto the next 
waveform display screen. 
 
  Waveform Display Screen 
ECG waveform (1CH when using 3-electrode lead cable, 2CH when using 
other lead cable), heart rate, pacemaker marker, respirogram, respiration rate, 
pulse wave, SpO
2
 measurement value, remaining battery level, and various 
messages are displayed. 
 
CAUTION 
  The LX-7230 does not have a diagnostic function. Check the diagnostic 
function on the receiving monitor. 
  The LX-7230 does not have an alarm function. Check the alarm function 
on the receiving monitor. 
  The ECG waveform size and sweep speed settings displayed on the 
LCD of the LX-7230 do not interface with the ones displayed on the 
screen of the receiving monitor. 
  The heart rate and respiration rate displayed on the LCD of the LX-7230 
may be different from the ones displayed on the receiving monitor. 
Because the algorithm of the ECG and respiration rate is different.