7. Measurement
Turn ON the power and the measurement starts.
■ Starting Screen
When the power is turned ON, the channel number configured on the LX-7230
is displayed at the top of the LCD.
is displayed at the top of the LCD.
Make sure whether the channel number on the
LCD matches the channel number indicated on
the label of the LX-7230 and the channel number
configured on the receiving monitor.
This screen automatically moves onto the next
waveform display screen.
LCD matches the channel number indicated on
the label of the LX-7230 and the channel number
configured on the receiving monitor.
This screen automatically moves onto the next
waveform display screen.
■ Waveform Display Screen
ECG waveform (1CH when using 3-electrode lead cable, 2CH when using
other lead cable), heart rate, pacemaker marker, respirogram, respiration rate,
pulse wave, SpO
other lead cable), heart rate, pacemaker marker, respirogram, respiration rate,
pulse wave, SpO
2
measurement value, remaining battery level, and various
messages are displayed.
CAUTION
The LX-7230 does not have a diagnostic function. Check the diagnostic
function on the receiving monitor.
The LX-7230 does not have an alarm function. Check the alarm function
on the receiving monitor.
The ECG waveform size and sweep speed settings displayed on the
LCD of the LX-7230 do not interface with the ones displayed on the
screen of the receiving monitor.
screen of the receiving monitor.
The heart rate and respiration rate displayed on the LCD of the LX-7230
may be different from the ones displayed on the receiving monitor.
Because the algorithm of the ECG and respiration rate is different.
Because the algorithm of the ECG and respiration rate is different.