Precautions about Waterproof 
CAUTION 
  Replace the “Battery Compartment Lid” of the LX-7230 regularly to keep 
the main unit performance of waterproof. If not regularly replaced, the 
quality of the lid will deteriorate and cannot keep the waterproof 
performance. For details about the regular replacement, contact your 
local Fukuda Denshi service representative. 
  The lid may be damaged from high impact. If the LX-7230 is dropped or 
is subjected to a high impact, make sure that the lid is not damaged. 
  However, the SpO
2
 probes (sensors/relay cables) are not waterproof. 
Do not take a bath with them, and ensure to be away from liquid. 
  Do not use the LX-7230 wet. Always wipe the LX-7230 with a soft cloth 
and dry it thoroughly before use. 
 
Precautions about ECG 
CAUTION 
  When removing electrodes from the patient, remove them carefully and 
slowly. Do not apply excessive force to remove them. Otherwise, it may 
damage the skin. 
  There are some cases when the pacemaker pulse cannot be detected 
depending on the pacemaker type, pulse voltage, pulse width, electrode 
lead type (unipolar, bipolar), electrode placement, or lead method which 
causes the pacemaker pulse amplitude to decrease and disables 
pacemaker pulse detection. 
  If signals similar to a pacemaker pulse are present, such as electric 
blanket noise or excessive AC frequency noise, these may be 
erroneously detected and displayed as a pacemaker pulse. In this case, 
check the condition of the electrodes and ECG lead cable to resolve the 
cause or turn off the pacemaker detection setting on the receiving 
monitor. 
vii