11. Troubleshooting
71
SpO
2
(LX-7230KM)
Message Cause
Solution
Probe is off.
Check the attached condition
of the probe.
of the probe.
Faulty Probe.
Replace the probe with new
one.
one.
Unspecified probe
is connected.
is connected.
Replace the probe with a
specified one.
specified one.
Check the attached condition
of the probe.
of the probe.
SpO
2
is not
measured correctly.
Cover the probe with an
opaque material to cut off
the outside light.
opaque material to cut off
the outside light.
When the pulse rate
is less than 30bpm
(under the
is less than 30bpm
(under the
measuring range
).
When the pulse rate
is more than
230bpm (
is more than
230bpm (
over the
measuring range
).
Due to excessive
body motion, SpO
body motion, SpO
2
is not measured
correctly
correctly
Keep the patient still.