Operator’s Manual  ZM-520PA/521PA/530PA/531PA
Connecting the Electrode Lead to the Transmitter
Connect the electrode lead to the ECG/RESP socket on the transmitter.
WARNING
After attaching the electrode to the patient and connecting the electrode lead to the 
transmitter, check that electrodes are attached to the patient and check that the 
electrode lead is connected to the transmitter properly. When the electrodes are 
removed from the patient, do not touch the metal part of the electrode with bare 
hands or let the metal part of the electrode contact the metal part of the bed or any 
other conductive parts. Failure to follow this warning may cause electrical shock or 
injury to the patient by discharged energy.
CAUTION
Do not shake or swing the transmitter while holding the leads or cables connected to 
the transmitter. The transmitter may come off and injure someone or damage 
surrounding instruments.
CAUTION
Hold the connector of the electrode lead when connecting/disconnecting the 
electrode lead. If you disconnect the electrode lead by pulling the lead, it damages 
the electrode lead.