Справочник Пользователя для Nihon Kohden Corporation ZM-520PA
84
Operator’s Manual ZM-520PA/521PA/530PA/531PA
5. Check that SpO
2
and pulse rate on the transmitter is within the following range.
SpO
2
on the SpO
2
Checker
Range
SpO
2
97%
95 to 99%SpO
2
(±2 digit)
80%
78 to 82%SpO
2
(±2 digit)
70%
67 to 73%SpO
2
(±3 digit)
Pulse rate
60 beats/min
57 to 62 beats/min (±3%/±1 beat/min)
120 beats/min
115 to 124 beats/min (±3%/±1 beat/min)
NOTE
The above range includes the error margin of the SpO
2
checker. Therefore, the range is
±1 digit outside the SpO
2
measuring accuracy of the transmitter in the “Specifications”
section.
6. Check that the SpO
2
and pulse waveform of the transmitter appear on the receiving monitor.