Operator’s Manual ZM-520PA/521PA/530PA/531PA
29
ECG MEASUREMENT
Turn ECG monitoring on or off. When electrodes are attached to the patient and ECG leads are
connected, ECG monitoring starts even when this setting is set to OFF.
Turn ECG monitoring on or off. When electrodes are attached to the patient and ECG leads are
connected, ECG monitoring starts even when this setting is set to OFF.
If this setting is set to OFF, the same setting on the receiving monitor must also be set to OFF.
NOTE
When “PROTOCOL” on the transmitter is set to 51 and the receiving monitor is able to
receive protocol 51, ECG measurement on the receiving monitor is automatically set
to OFF when this setting is set to OFF on the transmitter. To use protocol 51, the ORG-
9100A/ORG-9110A multiple patient receiver software version 03-03 or later is required.
receive protocol 51, ECG measurement on the receiving monitor is automatically set
to OFF when this setting is set to OFF on the transmitter. To use protocol 51, the ORG-
9100A/ORG-9110A multiple patient receiver software version 03-03 or later is required.
1. On the PARAMETER SETUP screen, press the
▼ key to move the cursor to “ECG
MEASUREMENT”.
2. Press the Function key.
3. Press
the
▼ key to select “ON” or “OFF”.
4. Press the Function key to register the selected setting. The cursor returns to “ECG
MEASUREMENT”.
RESP MEASUREMENT
Turn respiration monitoring on or off. When this setting is set to OFF, the same setting on the
receiving monitor is automatically set to OFF.
Turn respiration monitoring on or off. When this setting is set to OFF, the same setting on the
receiving monitor is automatically set to OFF.
1. On the PARAMETER SETUP screen, press the
▼ key to move the cursor to “RESP
MEASUREMENT”.
2. Press the Function key.
3. Press
the
▼ key to select “ON” or “OFF”.
4. Press the Function key to register the selected setting. The cursor returns to “RESP
MEASUREMENT”.