Operator’s Manual  ZM-520PA/521PA/530PA/531PA
 93
SpO
2
= 97
60
70
80
90
100
110
120
130
140
150
0
10
20
30
40
50
60
70
80
90
100
110
120
Time (seconds)
R
 (bpm)
PR Reference
PR
Transmitter
FCC regulation: 
FCC part 95 Subpart H
 
Wireless Medical Telemetry Service (WMTS)
Field strength limits: 
≤ 200 mV/m (at 3 m)
Undesired emissions: 
≤ 960 MHz: ≤ 200 μV/m (at 3 m)
 
≥ 960 MHz: ≤ 500 μV/m (at 3 m)
Antenna: Internal
Transmission channel: 
Indicated on the transmitter
Transmission frequency range: 
608.0250 to 613.9750 MHz
Channel spacing: 
50 kHz or 37.5 kHz (12.5 kHz when interleaved)
Modulation: 
FSK (frequency shift keying)
Type of emission: 
F1D
Occupied bandwidth: 
12 to 18 kHz
Effective radiated power: 
1.0 mW
Power Requirements
Rated voltage: 
3.0 V
Shutdown voltage: 
1.4 to 1.75 V
Operating voltage: 
Shutdown voltage to 3.2 V
Battery type: 
Two AA (LR6) type alkaline dry cell primary batteries
Battery lifetime (with alkaline batteries, at room temperature):
 
ZM-520PA: 
approximately 3.5 day
 
ZM-530PA: 
approximately 2.5 day (measuring ECG, respiration and  
 SpO
2
)
 
 
approximately 3.5 day (measuring ECG and respiration)