Справочник Пользователя для Nihon Kohden Corporation ZM-940PA
76
Operator's Manual ZM-940PA
3. LCD Display
Check that there are no dots missing on the LCD.
Check that there are no dots missing on the LCD.
1.
Remove one battery.
2.
While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
The SETUP screen appears.
3.
Press the NIBP INTERVAL key twice to move the cursor to “MANUAL CHECK”.
4.
Press the NIBP START/STOP key to
enter the MANUAL CHECK screen.
enter the MANUAL CHECK screen.
5.
Check that the cursor is on “LCD
TEST” and press the NIBP START/
STOP key.
TEST” and press the NIBP START/
STOP key.