Precautions 
 
1.
 
To place the product in front of the diaper, do the 
following two tests first to ensure that the urine is 
properly sensed: 
 
 
    (1)
 
Click the launcher on the black button, launcher LED 
lights will light up, while the alarm will also sound and 
light, which means launcher and alarm are all with 
power. 
 
    (2)
 
Dropping water in the sensing cable, the alarm will 
immediately sound and light, which means that the 
sensor cable and launcher have the correct link. 
    2.
 
Will be exposed to the skin of the spring clip place a 
sterile gauze to avoid direct skin contact. 
 
    3.
 
Sensory cable and spring clip should be placed along the 
groin below the urethra in the position near the 
perineum, male and female positions are different, 
please put appropriate position discretion, to avoid 
placing the sensor cable position or distance is too far 
can not be caused by sensing .       
               
 
    4.
 
To take off the launcher and sense the cable, please open 
the diaper or take off the diaper before removing the 
cable and launcher. 
 
    5. This product is for personal hygiene and to avoid 
cross-infection, sensing cable, rubber wrap and spring 
clip for personal use only, please do not for different 
users. 
 
    6. Do not reuse the sensing cable, rubber wrap and spring 
clip to avoid infection.   
 
7.
 
When the sensing cable is unplugged from the launcher, 
remove the sensing cable manually according to the 
socket end. Do not pull the cable part directly. Otherwise,