Troubleshooting 
 
This chapter summarizes the most common problems you could encounter with the LGT-232 
Portable Electro-Stimulation Therapy Device.  If you are unable to solve the problem with the 
information below, please call the distributors. 
Troubles 
Possible causes 
Solutions 
1. LCD Display does not 
light up at all when press 
the ON/OFF button 
The battery of the unit is empty. 
Charge the unit (see chapter 
Preparing for use’). 
2. The MStim Sport APP 
on the mobile phone 
could not connect with 
the unit. 
1. The unit is not turn on; 
2. Mobile phone Bluetooth is not 
open or mobile phone Bluetooth 
problems 
3. the distance  of the mobile 
phone and the unit is too far; 
4. The unit has contacted 
another mobile. 
The unit is asleep and Bluetooth 
is stopped. 
1. Turn on the unit; 
2. Please turn on the Bluetooth 
on the mobile phone or change 
another mobile phone; 
3. Please keep the mobile phone 
near the unit; 
4.  Disconnect the connected 
phone or restart the unit; 
5. Restart the unit. 
FOR-MRRD0001-EN232 UM0.1