Справочник Пользователя для Varex Imaging Deutschland AG XRPAD4336
w w w . p e r k i n e l m e r . c o m
X R p a d ™ 4 3 3 6 M E D 1 9
U
S
S
E
R
R
M
A
N
U
A
L
A
N
U
A
L
8.5.2
Interface and Power Unit XRpad™ IPU
The XRpad™ IPU is a Power Supply Unit with integrated additionally interfaces. The tethered
power and communication cable is connected to the XRpad™ 4336 MED detector. The
communication data are split inside the XRpad™ IPU into Gigabit Ethernet Interface, Detector
Trigger Interface, Hand Switch and Generator Interface and Detector Push Button Interface. The
Gigabit Ethernet Interface of the XRpad™ IPU is connected via Cat 5e/6 with the imaging
Workstation. The maximum cable length is 30m. The AC cable has to be connected to a properly
grounded receptacle. The AC cable is removable and will be plugged to an IEC connector. The XRD
IPU needs to be connected with a ground by the functional ground connector (Figure 6 (13)) or with
the potential of the hospital by the potential equalization connector (Figure 6 (12)) To isolate the
equipment electrically from supply mains on all poles simultaneously, the supply mains switch (Figure
6 (1)) must be used.
power and communication cable is connected to the XRpad™ 4336 MED detector. The
communication data are split inside the XRpad™ IPU into Gigabit Ethernet Interface, Detector
Trigger Interface, Hand Switch and Generator Interface and Detector Push Button Interface. The
Gigabit Ethernet Interface of the XRpad™ IPU is connected via Cat 5e/6 with the imaging
Workstation. The maximum cable length is 30m. The AC cable has to be connected to a properly
grounded receptacle. The AC cable is removable and will be plugged to an IEC connector. The XRD
IPU needs to be connected with a ground by the functional ground connector (Figure 6 (13)) or with
the potential of the hospital by the potential equalization connector (Figure 6 (12)) To isolate the
equipment electrically from supply mains on all poles simultaneously, the supply mains switch (Figure
6 (1)) must be used.
Figure 6 XRpad™ IPU Overview
Table 10 XRpad™ IPU Overview
Table 10 XRpad™ IPU Overview
1
XRpad™ IPU Power Switch
2
Power In
3
AC Input LED (Green) (
~
)
4
DC output LED
(Yellow) (
)
LED Yellow
DC Output ok, no output load
LED Green
DC Output ok, output loaded
5
XRpad™ Push Button ( )
6
XRpad™ Hand Switch In (Extension of the Push Button)
7
XRpad™ Interface and Power I/O
8
Trigger Out Signal to Generator
9
Trigger In Signal from Hand Switch (Prep / Expose)
10
Trigger In/Out I/F ( )
11
LAN port to Imaging Workstation
12
Potential Equalization Connector
13
Functional Ground Connector
WARNING
Connection of the XRpad™ IPU LAN port directly with the clinical
computer network may disturb the IT environment.
computer network may disturb the IT environment.
1 2
3 4 5
6
8 9 10 13
7
11 12