CAUTION: Do not touch the lamp with bare hands.  Dust, oils and grime will be transmitted to the 
lamp, and during use may cause the lamp to explode. 
 
CAUTION:  Lamp must be replaced after 1000 hours of use.   
 
NOTE: See Performance Characteristics Section for description of optical radiation emitted.  
 
3.5.7 Cleaning and Disinfection 
 
CAUTION: LuViva requires cleaning and disinfection between patients, as specified in this manual, 
to prevent operator-to-patient, and patient-to-patient contamination and cross-contamination. 
 
CAUTION: During cleaning and disinfection, wear personal protective equipment such as eye wear, 
moisture resistant clothing and chemical resistant gloves so that you are not exposed to chemical 
disinfectants and potentially infectious patient debris. Wear appropriate personal protective 
equipment. Always dispose of contaminated personal protective equipment as medical waste. 
 
CAUTION: Do not spray cleaning and disinfectant solutions directly into the rear and sides of the 
monitor, where the vent holes are located. 
 
CAUTION:  Use of cleaners other than those specified in Section 11 will void the warranty. 
 
CAUTION: Guided Therapeutics has performed cleaning validation and disinfection validation only 
for the class of cleaning agents and disinfectants listed in the procedure.  
 
CAUTION: If the HHU Window is not cleaned thoroughly and completely (with no debris or 
streaking), LuViva will not be able to accurately determine the disease state of the patient. 
 
 
3.5.8 Electrical Safety 
WARNING: No modification of this equipment is allowed. Modifications of the unit will void the 
warranty and may expose operators and patients to serious injury. 
 
CAUTION: To avoid the risk of electrical shock, this equipment must only be connected to a Supply 
Mains with protective earth (grounded wall outlet).Improperly powering LuViva may result in 
damage to LuViva and substantial risks of electric shock to the operator or patient. 
  
 
CAUTION: Risk of Electrical shock. Disconnect main power before replacing the lamp assembly or 
air filter. 
 
CAUTION:   Risk of Electrical Shock. No user-serviceable parts inside. Only components listed as 
“Replacement Parts” are replaceable by operator or service personnel. Refer servicing to qualified 
personnel. 
 
CAUTION: Position LuViva so it may be unplugged. 
 
CAUTION: Equipment not suitable for use in the presence of flammable anesthetic mixture with air, 
oxygen or nitrous oxide.