7.6 Dispose of Calibration Cap 
NOTE: Do not remove or discard the Calibration Cap until calibration has completed. 
 
 
Once calibration is complete, discard 
Cervical Guide Cap. (Recyclable ABS) 
 
7.7 Clean Cervix and Insert CG 
Clean the patient’s cervix with a swab to remove any excess mucus or blood. 
 
 
WARNING: Do not pinch vaginal or cervical tissue between the edges of the speculum blades 
and the CG.  Patient injury and pain may occur. 
 
CAUTION:  Prepare the patient for CG insertion and instruct the patient to be as still as possible to 
minimize potential injury during the test.  Unintended motion by the patient may cause an error, and 
the test will need to be repeated. 
CAUTION:   LuViva results depend upon the os of the cervix remaining in the center and the cervix 
remaining in the same orientation during the test sequence.  Movement by the patient may cause an 
error and the test will need to be repeated.