20.3.2 Install the Lamp Assembly 
CAUTION:  Wear long sleeves, gloves and face shield while installing the lamp in LuViva. 
CAUTION:  Do not drop the lamp, subject it to impacts, apply excessive force or scratch it:  the 
pressure inside the glass lamp is very high and might cause it to rupture.  Injury may result if the 
lamp is broken. 
CAUTION:  Never touch the lamp with bare hands.  Dust, oils and grime will be transmitted to the 
lamp, and during use may cause the lamp to explode. 
 
 
Open the Bulb Access Door 
of the Base Unit by 
unscrewing the locking 
screw with supplied 5/32 
Allen Key, and opening the 
hinged door. 
 
 
Open the Service kit and 
remove the Lamp 
Assembly. 
 
 
 Lamp Assembly enclosed 
in Placement Tool: Side 
View  
 
 
Lamp Assembly enclosed 
in Placement Tool: Top 
View