11.6 Software Upgrade
If software upgrades are needed, the upgrade software will be made available via USB drive. Install as per set-up
instructions.
1. Obtain the software upgrade USB drive.
2. Insert the software upgrade USB drive into the
USB port of the monitor.
3. Power up the system using Standby button on the
HHU.
4. After system startup, the GT Setup software will
install LuViva software.
5. The GT Setup screen will appear and display
progress.
6. At the end of the installation, shut down the
system.
7. Remove the setup USB.
The system is ready for use.
12. Service
The lamp assembly and air filter are the only operator serviceable components. These can be accessed from the lamp
access door with supplied tool.
There are no other operator serviceable parts. Do not attempt to open any other part of LuViva. Doing so could
damage LuViva, expose the operator to risk of electric shock and invalidate the manufacturer’s warranty.
access door with supplied tool.
There are no other operator serviceable parts. Do not attempt to open any other part of LuViva. Doing so could
damage LuViva, expose the operator to risk of electric shock and invalidate the manufacturer’s warranty.
WARNING: To avoid the risk of electrical shock, this equipment must only be connected to a
supply mains with protective earth (grounded wall outlet). Improperly powering LuViva may
supply mains with protective earth (grounded wall outlet). Improperly powering LuViva may
result in damage to LuViva and substantial risks of electric shock to the operator or patient.
CAUTION: Risk of Electrical shock. Disconnect main power before replacing the lamp assembly or air filter.
CAUTION: Risk of Electrical Shock. No user-serviceable parts inside. Only components listed as “Replacement Parts”
CAUTION: Risk of Electrical Shock. No user-serviceable parts inside. Only components listed as “Replacement Parts”
are replaceable by operator or service personnel. Refer servicing to qualified personnel.
13. Disassembly and Packing Instructions (for Service)
The Instrumentation package, which includes the Base Unit enclosure, umbilical cable and HHU can be completely
removed from the cart and returned to Guided Therapeutics, Inc. for service and/or repair as needed. A reusable
service case and additional packing instructions will be sent to transport the Instrumentation package.
If new packaging is needed to ship the device locally, please contact distributor and/or Guided Therapeutics, Inc. for
If new packaging is needed to ship the device locally, please contact distributor and/or Guided Therapeutics, Inc. for
return packaging and packing instructions.