Справочник Пользователя для Shenzhen Belter Health Measurement and Analysis Technology Co. Ltd. JK1V
78
mmHg
Pressure Accuracy
+/- 5 mmHg
Absolute accuracy of
pressure measurement
pressure measurement
Pulse Rate Range
40 - 240 BPM
Pulse Rate Resolution
1BMP
Cuff Regulation
+/- 10 mmHg
Cuff pressure will be
maintained to specified
pressure with this
tolerance
maintained to specified
pressure with this
tolerance
Min Pumping Rate
6 LPM
Minimum
rate
of
pumping air into cuff at
no load
no load
Power Input Voltage
24 V.D.C.+/- 5%
Regulated
24
VDC
power supply voltage
Max Power Input
Voltage Ripple
Voltage Ripple
100mv pk-pk
Max Occlusion
Duration
Duration
20 minutes
The maximum duration
that either of the cuffs
can be inflated
that either of the cuffs
can be inflated
Battery voltage
D.C. 14.8V
The
built-in
battery
lithium battery
Product Labels