When holding the【M】memory button the user can view the data from the most recent
date to the oldest date.
d) If the data in memory displays the heart mark, it prompts when measurement the
irregular pulse is detected.
e)
If the user need to display the other set of memory data please refer to section 3.2.2
Caution: Continuous holding the 【M】button will delete all the memory.
2)
Delete Memory Data
a)
Enter into the memory mode refer to section 3.2.2.
b) Press and hold the【M】and【ON/OFF】button until the “---” displayed.
c)
The unit will only delete the present set of memory data; the other set of memory data
will not be affected.
d) The device is not capable to delete a single data.
e)
Press the【ON/OFF】button for 3 seconds to exit the memory mode and turn off.
5. Error Indication
List of error codes are as follows:
Error
Cause
How to correct
Cuff
self-test
symbol
Power on, cuff inflation rate is too
low or main unit does not connect
with the cuff or cuff is too loose.
low or main unit does not connect
with the cuff or cuff is too loose.
(1)
Cuff or bladder leakage, please connect
manufacturer.
manufacturer.
(2)
Confirm the cuff is wrapped up
correctly (ref 4.2), retake the
measurement.
correctly (ref 4.2), retake the
measurement.
Er 2
Weak signal or cuff is too loose.
Cuff too loose, Confirm the cuff is wrapped up
correctly (ref 4.2) retake the measurement.
correctly (ref 4.2) retake the measurement.
Er 3
Heavy shock
during the
measurement.
The fingers do not force when measurement.
The measurement wrist do not bent up or
down, do not clenched fist. Then retake the
measurement (ref 4.3).
The measurement wrist do not bent up or
down, do not clenched fist. Then retake the
measurement (ref 4.3).
Er 5
Bad signal, moving or talking
during the measurement.
during the measurement.
Remain still, and retake the measurement (ref
4.1).
4.1).
Er7
Measurement abnormal.
Please retake the measurement.
Lo
Low battery power, cannot inflate.
Change the batteries (ref 3.1).
6. Trouble Shooting
When the unit encounters malfunction during the use, refer to table below:
Abnormal
How to correct
After batteries installation, power
on, no display.
on, no display.
(1) Check batteries polarity.
(2) If still cannot power on, reinstall the batteries or
(2) If still cannot power on, reinstall the batteries or
change new batteries.
Measured value is abnormally high (1) Confirm the cuff is wrapped up correctly.