Справочник Пользователя для SHENZHEN PUMP MEDICAL SYSTEM CO. LTD. BFW01

Скачать
Страница из 17
                 
                                                       
  Rev. A1
 
 
c) 
When holding the【M】memory button the user can view the data from the most recent 
date to the oldest date. 
d)  If the data in memory displays the heart mark,  it prompts when measurement the 
irregular pulse is detected. 
e) 
If the user need to display the other set of memory data please refer to section 3.2.2 
Caution: Continuous holding the  【M】button will delete all the memory. 
 
2) 
Delete Memory Data   
a) 
Enter into the memory mode refer to section 3.2.2. 
b)  Press and hold the【M】and【ON/OFF】button until the “---” displayed. 
c) 
The unit will only delete the present set of memory data; the other set of memory data 
will not be affected. 
d)    The device is not capable to delete a single data. 
e) 
Press the【ON/OFF】button for 3 seconds to exit the memory mode and turn off. 
 
5.    Error Indication 
List of error codes are as follows: 
Error 
Cause 
How to correct 
Cuff 
self-test 
symbol 
Power on, cuff inflation rate is too 
low or main unit  does not connect 
with the cuff or cuff is too loose. 
(1) 
Cuff or bladder leakage,  please connect 
manufacturer. 
(2) 
Confirm the cuff is wrapped up 
correctly (ref 4.2), retake the 
measurement. 
Er 2 
Weak signal or cuff is too loose. 
Cuff too loose, Confirm the cuff is wrapped up 
correctly (ref 4.2) retake the measurement. 
Er 3 
Heavy shock 
during the 
measurement. 
The fingers do not force when measurement. 
The measurement wrist  do not bent up or 
down, do not clenched fist. Then  retake the 
measurement (ref 4.3). 
Er 5 
Bad  signal,  moving or talking 
during the measurement. 
Remain still, and  retake the measurement  (ref 
4.1). 
Er7 
Measurement abnormal. 
Please retake the measurement. 
Lo 
Low battery power, cannot inflate. 
Change the batteries (ref 3.1). 
 
 
6.    Trouble Shooting 
When the unit encounters malfunction during the use, refer to table below: 
Abnormal 
How to correct 
After batteries installation, power 
on, no display. 
(1)  Check batteries polarity. 
(2)  If still cannot power on, reinstall the batteries or 
change new batteries. 
Measured value is abnormally high  (1)  Confirm the cuff is wrapped up correctly.