KO00065K
VC150 Vital Signs Monitor
8-21
Masimo SpO
2
Masimo rainbow
®
 SET
®
 SpO
2
 and special features
WARNINGS
Variation in hemoglobin measurements may be profound and may be 
affected by sample type, body positioning, as well as other physiological 
conditions. As with most hemoglobin tests, VC150 test results should be 
scrutinized in light of a specific patient’s condition. Any results 
exhibiting inconsistency with the patient’s clinical status should be 
repeated and/or supplemented with additional test data.
Inaccurate SpHb and SpOC readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue
Externally applied coloring and texture such as nail polish, acrylic 
nails, glitter, etc.
Elevated levels of bilirubin
Low arterial perfusion
Motion  artifact
Low arterial oxygen saturation levels
Elevated carboxyhemoglobin levels
Elevated methemoglobin levels
Difference between patient's finger skin and finger core 
temperature
Hemoglobin synthesis disorders
Hemoglobinopathies and synthesis disorders such as thalassemias, 
Hb s, Hb c, sickle cell, etc.
Vasospastic disease such as Raynaud's
Elevated altitude
Peripheral vascular disease
Liver disease
EMI radiation interference
Inaccurate SpCO and SpMet readings may be caused by:
Intravascular dyes such as indocyanine green or methylene blue
Abnormal hemoglobin levels
Low arterial perfusion
Low arterial oxygen saturation levels including altitude induced 
hypoxemia
Elevated total bilirubin levels
Motion  artifact
SpCO readings may not be provided if SpO
2
 readings are less than 
90%
SpCO readings may not be provided if SpMet readings are greater 
than 2%