10-2
VC150 Vital Signs Monitor
KO00065K
Pulse rate: Description
Description
The PR (pulse rate) measurement displays heartbeats per minute (bpm), the 
source of the heartbeat data, a blinking heart icon and upper and lower alarm 
limits for the parameter. A beep provides an audible representation of the 
heartbeat. Also, the heart icon and the beep follow suit with the actual rhythm of 
the heart. The faster the heartbeat, the faster the icon blinks and the beep 
sounds, and vice versa.
The PR parameter receives data and waveform either from the NIBP parameter 
or the SpO
2
 parameter (Masimo SET
®
, Nellcor or GE TruSignal). The SpO
2
 
parameter is always the primary data source for PR (displayed in yellow) 
whereas NIBP is the secondary (the PR displayed in the same color as the NIBP). 
If pulse rate data cannot be obtained from SpO
2
, two dashes ‘- - ’ are displayed 
in the PR window. If there is a recent enough NIBP for PR, it is displayed. The data 
source is also displayed in technical area underneath the parameter value. Refer 
to the individual SpO
2
 and NIBP chapters for more details on NIBP and SpO
2
.
When the NIBP measurement is complete, a value is displayed in the PR window. 
The value is displayed as long as the results of that determination are displayed 
or until SpO
2
 switches to monitoring mode.
NOTES
If Masimo rainbow SET
®
 is the data source, the pulse rate values are filtered 
by an averaging technique that determines how quickly the reported values 
respond to changes in the patient’s saturation. Increased averaging time 
affects time to alarm for SpO
2
 saturation and pulse rate limits.
When NIBP is in STAT mode and is the source of pulse rate, the pulse rate 
value is not checked against its limits upon completion of the measurement.
Due to the algorithms the various sources use to measure the heartbeat, 
values in the PR window may differ when the monitor changes from one 
source to another.
A patient’s vital signs may vary dramatically during the use of 
cardiovascular agents such as those that raise or lower blood pressure or 
those that increase or decrease heart rate.
If an SpO
2
-derived pulse rate is erratic, the pulse oximeter parameter may 
be unable to measure the pulse. If the pulse beep tone does not sound with 
each pulse, the pulse beep volume is turned off, or the speaker is 
malfunctioning.
Pulse rate alarm limits
Refer to 
on page 3-14 to set uUpper and Lower alarm limits 
for Pulse Rate. The range for all sources is the same: Upper is 35 to 235 bpm and 
Lower is 30 to 230 bpm. The limits can be set in increments of 5 bpm.
When NIBP is the source for STAT mode measurement, the pulse rate value is 
not checked against the limits.