Справочник Пользователя для Innokas Yhtyma Oy VC150

Скачать
Страница из 332
KO00065K
VC150 Vital Signs Monitor
12-7
Exergen temperature: Exergen temperature specifications
Exergen temperature specifications
Exergen scanner battery specifications
Units of measure
°Celsius (C) or °Fahrenheit (F). Defined by purchased Exergen scanner.
Range
Measurement mode
Max: 43° C (110° F)
Min.: 16° C (61° F)
Accuracy
±0.1° C (±0.2° F)
complies with EN 12470-5
Predictive mode
Not applicable
Monitoring mode
Not applicable
Operating environment (ambient)
16° to 40° C (61° to 104° F)
Arterial head balance range for body 
temperature
1
 
34.5° to 43° C (94° to 110° F)
Resolution 
0.1° C or 0.1° F
Response time
Approximately 0.04 seconds, typical
Calibration
Oral or Core
1
 Automatically applied when temperature is within normal body temperature range, otherwise reads surface 
temperature.
NOTE
Use only Exergen probe covers. The size and shape of the probe covers can affect the performance of the 
scanner. Inaccurate readings occur unless the proper probe covers are used. Refer to the VC150 supplies and 
accessories document for reorder part numbers.
Capacity
One 9 volt alkaline
Battery life
Approx. 15,000 readings
(When scanning for 5 seconds and reading the temperature display for 2 
seconds before turning thermometer off.)