VC150 Vital Signs Monitor
B-9
Maintenance: Repairs
Repairs
If your product requires warranty, extended warranty, or non-warranty repair 
service, contact Innokas Medical Technical Support or your local Innokas 
Medical representative. 
Estimates for non-warranty repairs are provided at no charge; however, the 
product must be sent to Innokas Medical for an estimate. To facilitate prompt 
service in cases where the product has external chassis or case damage, please 
advise the representative when you call. 
The representative will record all necessary information and will provide a 
Return Authorization Number. Prior to returning any product for repair, a Return 
Authorization Number must be obtained.
Changing the Exergen temperature unit battery
Required tools and parts
Paper clip or metal pin of equal thickness
A new quality alkaline 9V battery
NOTE
If the Exergen scanner is not used regularly, remove the battery to prevent 
possible damage due to chemical leakage.
Procedure
1. Unplug the scanner cable from the monitor USB port.
2. Bend a paper clip or obtain a metal pin of equal thickness and then insert it 
into a small hole on the side of the Exergen frame (A).
3. Push the pin inside until the battery cover (B) is released.
4. Remove the cover.
5. Remove and disconnect the battery (C).
6. Replace the battery.
7. Place the lower end of the cover (D) in the groove.
8. Then push the upper part (E) down until it locks up.