Справочник Пользователя для Innokas Yhtyma Oy VC150
2-12
VC150 Vital Signs Monitor
KO00065K
Product overview: Operating modes
NOTES
•
There are no alarm limits and no clinical alarm messages for temperature
measurement.
measurement.
•
All changes are temporary and return to the default configuration settings
when the monitor is turned off or a new patient is admitted. To permanently
change the alarm settings refer to
when the monitor is turned off or a new patient is admitted. To permanently
change the alarm settings refer to
12.
WARNINGS
Monitors located in the same clinical area may contain different alarm
default settings, which can result in a potential hazard.
default settings, which can result in a potential hazard.
Always check your alarm settings before using the monitor.
Caution should be taken to not set alarm limits to extreme values, as
this can render the alarm system useless.
this can render the alarm system useless.
Monitor Setup
Monitor Setup consists of four tabs that all users are allowed to configure. These
configurations are only for the current patient. To make these changes
permanent, these must be changed in the configuration mode.
configurations are only for the current patient. To make these changes
permanent, these must be changed in the configuration mode.
on page 3-19 for more
information.
NIBP is used to select cuff position, target inflation pressures and patient
position. Refer to
position. Refer to
on page 6-1 for more information.
SpO
2
is used to select measurement site and visual elements of the SpO
2
parameter in clinical mode. Refer to
on page 9-1for more
information.
Temperature is used to configure temperature measurement. Refer to
on page 3-25 for more information.