Getting started: Setting up NIBP connections
CAUTION
The RADIAL-CUF has been validated for use only with GE SuperSTAT
algorithm for adult obese patients. The RADIAL-CUF has not been
validated for use against the GE Auscultatory algorithm. Refer to the
RADIAL-CUF instructions for use for sensor requirements.
algorithm for adult obese patients. The RADIAL-CUF has not been
validated for use against the GE Auscultatory algorithm. Refer to the
RADIAL-CUF instructions for use for sensor requirements.
3. Inspect cuff/adapter/air hose for damage. Replace cuff when aging, tearing,
leaks or weak closure is apparent. Do not inflate cuff when unwrapped.
CAUTIONS
Incorrect cleaning procedures will contaminate the accessory or
shorten lifetime of the cuffs or hoses. Clean according to the specific
cleaning and disinfection instructions provided with the accessory.
shorten lifetime of the cuffs or hoses. Clean according to the specific
cleaning and disinfection instructions provided with the accessory.
Do not use cuff if structural integrity is suspect.
4. Connect the cuff to the air hose. Refer to
on page 6-1 of this manual
for complete cuff connection instructions.
CAUTION
Always use the appropriate hose and cuff combination for the patient.
Any attempt to modify the hose may prevent the monitor from
switching between the neonate and adult/pediatric measurement
modes.
Any attempt to modify the hose may prevent the monitor from
switching between the neonate and adult/pediatric measurement
modes.
NOTE
Ensure the cuff and the hose are securely connected. Make sure no leak
occurs.
occurs.
5. Refer to
on page 6-1 of this manual for complete instructions on
taking an accurate NIBP determination.
NOTES
NOTES
•
Use only GE BP cuffs. The size, shape, and bladder characteristics can
affect the performance of the instrument. The accuracy of the
DINAMAP blood pressure algorithm has only been confirmed when used
with GE cuffs. Refer to the VC150 supplies and accessories document
for order part numbers.
affect the performance of the instrument. The accuracy of the
DINAMAP blood pressure algorithm has only been confirmed when used
with GE cuffs. Refer to the VC150 supplies and accessories document
for order part numbers.
•
When the monitor displays Adult in Monitor Setup > NIBP, it
encompasses both adult and pediatric patients.
encompasses both adult and pediatric patients.