KO00065K
VC150 Vital Signs Monitor
3-3
Getting started: Setting up NIBP connections
CAUTION
The RADIAL-CUF has been validated for use only with GE SuperSTAT 
algorithm for adult obese patients. The RADIAL-CUF has not been 
validated for use against the GE Auscultatory algorithm. Refer to the 
RADIAL-CUF instructions for use for sensor requirements.
3. Inspect cuff/adapter/air hose for damage. Replace cuff when aging, tearing, 
leaks or weak closure is apparent. Do not inflate cuff when unwrapped.
CAUTIONS
Incorrect cleaning procedures will contaminate the accessory or 
shorten lifetime of the cuffs or hoses. Clean according to the specific 
cleaning and disinfection instructions provided with the accessory.
Do not use cuff if structural integrity is suspect.
4. Connect the cuff to the air hose. Refer to 
on page 6-1 of this manual 
for complete cuff connection instructions.
CAUTION
Always use the appropriate hose and cuff combination for the patient. 
Any attempt to modify the hose may prevent the monitor from 
switching between the neonate and adult/pediatric measurement 
modes.
NOTE
Ensure the cuff and the hose are securely connected. Make sure no leak 
occurs. 
on page 6-1 of this manual for complete instructions on 
taking an accurate NIBP determination.
NOTES
Use only GE BP cuffs. The size, shape, and bladder characteristics can 
affect the performance of the instrument. The accuracy of the 
DINAMAP blood pressure algorithm has only been confirmed when used 
with GE cuffs. Refer to the VC150 supplies and accessories document 
for order part numbers.
When the monitor displays Adult in Monitor Setup > NIBP, it 
encompasses both adult and pediatric patients.