12-2
VC150 Vital Signs Monitor
KO00065K
Exergen temperature: Description
Description
The monitor can use the Exergen Temporalscanner technology if it has the
Exergen temperature parameter (temperature parameter displayed on the
screen and Exergen connected to the monitor USB port).
Exergen temperature parameter (temperature parameter displayed on the
screen and Exergen connected to the monitor USB port).
Temporalscanner technology
provides a method of temperature
assessment based on infrared
measurement of the thermal
radiation of the skin. The temporal
artery is used as a sampling site
because of its relatively constant
perfusion rate. While the scanner
senses temperature from the area
of the temporal artery, it can
optionally report temperature
readings referenced to either
central arterial temperature or oral
temperature.
provides a method of temperature
assessment based on infrared
measurement of the thermal
radiation of the skin. The temporal
artery is used as a sampling site
because of its relatively constant
perfusion rate. While the scanner
senses temperature from the area
of the temporal artery, it can
optionally report temperature
readings referenced to either
central arterial temperature or oral
temperature.
Temperature values are shown in
degrees Celsius or Fahrenheit on
the monitor screen and on the
scanner’s LED display screen. The
LED display screen displays the
temperature, but does not display
the unit of measurement.
degrees Celsius or Fahrenheit on
the monitor screen and on the
scanner’s LED display screen. The
LED display screen displays the
temperature, but does not display
the unit of measurement.
The preset unit of temperature measurement on the monitor screen and the
Exergen LED display can be changed by service personnel only.
Exergen LED display can be changed by service personnel only.
NOTES
Arterial temperature is close to rectal temperature, approximately 0.4° C
(0.8º F) higher than oral temps. Expect larger differences at times, however,
as the dynamics of thermoregulation favor the temporal artery method.
Arterial temperature is the same temperature as the blood flowing from the
heart via the pulmonary artery. It is the best determinant of body
temperature, and unaffected by the artifactual errors and time delays
present with oral and rectal methods.
(0.8º F) higher than oral temps. Expect larger differences at times, however,
as the dynamics of thermoregulation favor the temporal artery method.
Arterial temperature is the same temperature as the blood flowing from the
heart via the pulmonary artery. It is the best determinant of body
temperature, and unaffected by the artifactual errors and time delays
present with oral and rectal methods.
The scanner takes a single instance of a temperature measurement.
Exergen technology does not support continuous monitoring.
Exergen technology does not support continuous monitoring.
A monitor with the Exergen temperature technology configuration setting
cannot perform Welch Allyn temperature measurements and vice versa.
cannot perform Welch Allyn temperature measurements and vice versa.
Use this product only for its intended use as described in this manual.
Do not take temperature over scar tissue, open sores, or abrasions.
The thermometer is not shockproof. Do not drop it or expose it to electrical
shocks.
shocks.
Do not autoclave. Please note cleaning and sterilizing procedures in this
manual.
manual.