Справочник Пользователя для Contec Medical Systems Co. Ltd. CMS50D-BT
Figure 2 Sketch map for Pulse Oximeter
4.2 Interface introduction
4.3 Parameter introduction
4.3 Parameter introduction
Parameter Description
Unit
%SpO
2
Oxygen saturation
%
PRbpm
Pulse rate
bpm
Calorie
Calorie Calorie
Steps Step Step
5 Operation guide
5.1 Installation of batteries
Figure 5 Figure 6 Figure 7
1)
Open the battery compartment cover on the back of the device, shown as Figure 5.
2)
Install the two “AAA” Alkali batteries, pay attention to the polarity, shown as Figure 6..
3)
Install the battery compartment, shown as Figure 7.
Notice
Please make sure the batteries polarities are installed properly, to avoid any damage to the device.
Please take out the batteries if the device is not used for a long time.
Please replace the batteries in time when low-battery appears.
Please replace two new batteries with same type at the same time.
After replacing batteries, perform sync time with APP again.
Notice
Disposal of scrap batteries should follow local regulations about environment.
5.2 Data measurement
1)
5.2 Data measurement
1)
Install the two “AAA” batteries properly, then the device will turn on automatically and enter to
“Synchronous Time...” interface.
“Synchronous Time...” interface.
Figure 8 Synchronous time interface
Notice
In “Synchronous Time...” interface, Bluetooth will be turned on automatically.
2)
2)
Three methods to exit from “Synchronous Time...” interface.
Notice
Before each measurement, make sure the device has been performed sync time, to insure accurate time for
storing data. It indicates that the device needs to perform sync time when it displays “Synchronous
Time...”, otherwise it will not store measurement results.
The first method:
Don’t perform sync time, press the button, then the device will enter to “Steps” interface from “Synchronous
Time...” interface.
The second method:
Don’t perform sync time, wait for several minutes, then the device will enter to “Steps” interface from
“Synchronous Time...” interface.
The third method:
Perform sync time, connect to power, then the device will turn on automatically, when it enters to “Synchronous
Time...” interface, connect it with App, then it will automatically adjust time.
3)
storing data. It indicates that the device needs to perform sync time when it displays “Synchronous
Time...”, otherwise it will not store measurement results.
The first method:
Don’t perform sync time, press the button, then the device will enter to “Steps” interface from “Synchronous
Time...” interface.
The second method:
Don’t perform sync time, wait for several minutes, then the device will enter to “Steps” interface from
“Synchronous Time...” interface.
The third method:
Perform sync time, connect to power, then the device will turn on automatically, when it enters to “Synchronous
Time...” interface, connect it with App, then it will automatically adjust time.
3)
In “Steps” interface, insert the finger to the probe (fully insert), then the device will automatically enter to
“SpO
“SpO
2
” interface from “Steps” interface, shown as Figure 3.
Notice
If the time is wrong, the device will not store data.
Fingernails and the luminescent tube should be on the same side.
During using, the testee’s finger should not shake, and keep quiet.
4)
Read SpO
2
and pulse rate value directly from the screen after inserting the finger for about 4~8 s, shown as
Figure 3. Pull out the finger after stable data appears to finish data storing. Only one result is stored from
each inserting to pulling out the finger.
each inserting to pulling out the finger.
5)
In SpO
2
state, screen direction can be changed automatically.
6)
The interface between “Steps” and “SpO
2
” interface can be switched by inserting the finger.
a.
In “Steps” interface, insert the finger to enter to “SpO
2
” interface.
b.
In “SpO
2
” interface, pull out the finger to enter to “Steps” interface, shown as Figure
Notice
In “Steps” interface, Bluetooth will be in “ON” state all the time.
7)
7)
In “Steps” interface for 1 minute, it will automatically enter standby mode.
Notice
In standby mode, the screen does not display any contents, but the pedometer can work normally.
In standby mode, Bluetooth is in “OFF” state.
8)
In standby mode, Bluetooth is in “OFF” state.
8)
In standby mode, press Button to turn on screen display and enter to “Steps” interface shown as Figure 4.
5.3 Data upload
The device can transmit SpO
The device can transmit SpO
2
, pulse rate, Steps and calorie value to phone-side by wireless mode.
User can add the device by searching it in PC-side, as each device has the unique code which is marked in the
device enclosure, which provides convenience for user to upload the results to PC, avoids confusion caused by
excessive and same devices. After measuring, the device will automatically connect to the PC communicated with
it last time.
In “STEPS” interface, Bluetooth is in “ON” state, and it is in “OFF” state in other interfaces.
device enclosure, which provides convenience for user to upload the results to PC, avoids confusion caused by
excessive and same devices. After measuring, the device will automatically connect to the PC communicated with
it last time.
In “STEPS” interface, Bluetooth is in “ON” state, and it is in “OFF” state in other interfaces.
Notice
Before uploading steps data, please do sports for 5 minutes at least.
The Pedometer can only upload the data(measured in the last day) on the next day.
5.4 Accessories
A. A lanyard
B. Two batteries(optional)
C. A User Manual
D. A SpO
The Pedometer can only upload the data(measured in the last day) on the next day.
5.4 Accessories
A. A lanyard
B. Two batteries(optional)
C. A User Manual
D. A SpO
2
package
6 Maintenance
6.1 Routine maintenance
6.1.1 Cleaning and disinfecting
Turn off the device before cleaning. Use medical ethanol to clean and disinfect the device, natural air-drying or
clean it with a clean and dry cloth. Avoid any liquid entering to the device.
6.1.2 Maintenance
1)
6.1.1 Cleaning and disinfecting
Turn off the device before cleaning. Use medical ethanol to clean and disinfect the device, natural air-drying or
clean it with a clean and dry cloth. Avoid any liquid entering to the device.
6.1.2 Maintenance
1)
Keep the device away from dust, high temperature and humidity, and avoid strong shaking and hitting.
2)
When the device needs to clean, please wipe its surface with neutral detergent, avoid using strong corrosive
liquid, like alcohol and gasoline, etc.
liquid, like alcohol and gasoline, etc.
3)
Please clean and disinfect the device before using according to the User Manual (6.1.1).
4)
Please replace the batteries in time when low-battery appears.
5)
Please take out the batteries if the device is not used for a long time.
6)
The device needs to be calibrated periodically (or according to the calibrating program of hospital). It also
can be performed at the state-appointed agent or just contact us for calibration.
can be performed at the state-appointed agent or just contact us for calibration.
6.2 Transportation and storage
A. The packed device can be transported by ordinary conveyance or according to transport contract. The device
A. The packed device can be transported by ordinary conveyance or according to transport contract. The device
can not be transported mixed with toxic, harmful, corrosive material.
B. The packed device should be stored in room with no corrosive gases and good ventilation. Temperature:
-40 °C~60 °C; Relative Humidity: ≤95 %; Atmospheric pressure: 500 hPa~1060 hPa
7 Troubleshooting
Trouble
Analysis of cause
Solution
The SpO
2
or
Pulse Rate can
not be
displayed
normally
not be
displayed
normally
1) The finger is not properly inserted.
2) The SpO
2) The SpO
2
value of patient is too low to be
detected.
1) Insert the finger properly and try
again.
2) Please try several times more; Go to a
hospital for a diagnosis if you are sure
the device works all right.
the device works all right.
The SpO
2
or
Pulse Rate are
not displayed
stably
not displayed
stably
1) The finger is not inserted inside deep
enough.
2) The finger is shaking or the patient is
moving.
1) Insert the finger properly and try
again.
2) Please do not move.
The device can
not be turned
on
not be turned
on
1) The batteries are drained or almost
drained.
2) The batteries are installed improperly.
3) The device’s malfunction.
3) The device’s malfunction.
1) Please replace the batteries.
2) Please install the batteries again.
3) Please contact the local service center.
2) Please install the batteries again.
3) Please contact the local service center.
The display
disappears
suddenly
disappears
suddenly
1) Low voltage
2) The device enters to standby mode
2) The device enters to standby mode
automatically if there is no operation for 1
minute.
minute.
1) Please change the batteries.
2) Normal.
2) Normal.
Data can not be
stored.
stored.
1) False time.
2) Measurement time is too short.
2) Measurement time is too short.
1) Sync time.
2) Normal.
2) Normal.
8 Meaning of Symbol
Symbol
Meaning
Warning – refer to User Manual
Refer to instruction manual/booklet
%SpO
2
The pulse oxygen saturation(%)
PRbpm Pulse
rate(bpm)
Calorie Calorie
Low-voltage
Battery anode
Battery cathode
IP22
Ingress of liquids rank
Type BF applied part
9 Specification
Display Information
Display Mode
The Pulse Oxygen Saturation(SpO
2
)
2-digit digital OLED display
Pulse Rate(PR)
3-digit digital OLED display
Pulse Intensity(bar-graph)
bar-graph OLED display
Calorie
5-digit digital OLED display
Steps
5-digit digital OLED display
Time
Year-month-day hour-minute OLED display
SpO
2
parameter
Measurement range
0 %~99 % (resolution: 1 %)
Pulse parameter
Measurement range
Measurement range
30 bpm~250 bpm (resolution: 1 bpm)
Pedometer
Measurement range
Measurement range
0~65535 steps (resolution: one step)
Safety classification
Internally powered equipment, type BF applied part
Pulse Intensity
Range
Continuous bar-graph display, the higher display indicates the
stronger pulse.
stronger pulse.
Power supply
Two “AAA” batteries
Two “AAA” batteries
Dimension and weight
Dimension
58 mm(L) x 32 mm(W) x 34 mm(H)
Weight
About 52 g (including batteries)