Intended for use
Ecare IA
The 
Blood Glucose Monitoring Systems is intended for the 
quantitative measurement of glucose in fresh venous whole blood, 
or in fresh capillary whole blood from the fingertip. It is intended people 
with diabetes mellitus at home or healthcare facilities for monitoring 
the effectiveness of diabetes control program. It is not intended for the 
diagnosis of or screening for diabetes mellitus, and is not intended for 
use on neonates.
Test Principle
The 
 blood glucose monitoring system is plasma calibrated 
Ecare IA
to allow easy comparison of results with laboratory methods. Glucose in 
the blood sample mixes with special chemicals on the test strip and a 
small electrical current is produced. This current is measured by the 
Meter and displayed as your blood glucose result. The strength 
Ecare IA 
of these currents changes with the amount of glucose in the blood sample.
Ecare IA Blood Glucose Test strip
Note
1.
Meter should only be used with 
 Blood Glucose Test Strip. Using 
Ecare IA 
Ecare IA
    other test strips with this meter can produce inaccurate results.
2.The front side of the 
Blood Glucose Test Strip should face up when inserting 
Ecare IA 
    the 
 Blood Glucose Test Strip.
Ecare IA
3.The strips are for single use only, please do not reuse or use the used strip and repeat the test.
Please insert the test strip               The temperature is out of range
Please apply blood to the strip       Control solution mode
                                          Measurement Unit
                                                                                  
 Milligram per Deciliter
 Millimole per Liter
1. Ecare IA  Meter × 1  .Lancet device × 1                         3. Ecare IA Blood Glucose
                                                                                                                Test Strip vial × 1
4. Lancet × 10        5.Ecare GDH Control Material      6.AAA  Alkaline Battery  
                                 
                                    (Optional)
Protective cover
Lancet point
Operation of Lancing device
• Alternative site testing means that people test blood glucose from the forearm,
  upper arm or finger tip.
• Please consult your doctor to determine if alternative site testing is right for
  you.
Select a soft, fleshy area of skin that is free from hair, moles and visible 
• 
  veins for alternative site testing.
Avoid lancing areas with obvious veins or moles to avoid excess bleeding 
• 
  or where tendons or bones stick out (for example, hand bones and ankle 
  bones)
AST alternate site testing
It is strongly recommend that do AST ONLY in the following intervals
-In a pre-meal or fasting state (more than 2 hours since the last meal).
-Two hours or more after taking insulin.
-Two hours or more after exercise.
Do Not do AST
-You think your blood glucose is low.
-You are unaware of hypoglycemia.
-Your AST results do not match how you feel.
-You are testing for hyperglycemia.
-Your routine glucose results are often fluctuating.
-If you are pregnant.
  Run a control test by Ecare GDH Control Material (optional accessory)
  when you open a new vial of test strips or if you think that your test 
  result is incorrect. Running a control test lets you know that the meter
  and test  strips are working properly.
IA Blood Glucose
Monitoring System
User’s Manual
2    AAA
            
1.Press S button 2 seconds into the setting mode.
    (no test strip inserted).   
    Please follow the below steps to change the 
    parameters.
2.The “L-1”  on the display.
    Press "M" to select the languages(L-1~L-3) 
    ,then press "S" to confirm language:
3.The “3”  on the display. 
    Press "M" to select the volume(3~0) ,
    then press "S" to confirm volume:    
    3:high volume
    0:mute
1.The year flashes on the display.
    Press "M" to select the year(2015~2050) 
    ,then press "S" to confirm year.
2.Repeat the same steps to set
   the month, date ,hour and minute.
Note
The hour format is the 
12-hour time.
Memory
symbol
Control
Solution
Mode
Set the measurement unit 
                   
1.The measurement unit flashes.
2.Set the measurement unit:
   Press  "M"to select the unit :mg/dL or mmol/L.
3.Press "S" to confirm measurement unit.
    
    
Note
     1. The unit can be changed in setting mode.
     2. If you can not change unit during step 2, it means the
         system was fixed the unit by default setting.
     
              
  
   
1. Wash hands with soap and warm water. 
     Rinse and dry thoroughly.
2. Pull out a test strip from the box and 
    must re-cap the box immediately.
3. Insert the test strip, the front side of the 
    test strip face up, into the test port. 
    The meter will  automatically turn on.
  
1. The “dEL” display, press “S” goes to sleep    
    mode.   
2. Or press “M” 2 sec, it displays “OK” 
     to delete all memory, then goes to
     sleep mode. 
Delete the test result
Preparing your blood sample
1. Wash your hands with soap and warm water.
    Then rinse and dry thoroughly.
2. Massage the puncture site.
3. Using lancet device firmly on the side of the fingertip
     to get a drop of blood.
4. Press the button on the lancet device.
    The blood has been taking as you hear a click sound.
5. Gently squeeze or massage your finger until drop of
     blood forms.
Note
• The user must wash hands thoroughly with soap and water after handling
   the meter, lancet device or test strips.
• The lancets can not to be re-used or used more than one person to increase
   the infection risk.
• Don’t use Povidone  Iodine Solution or any org nic solvent for cleaning!
• Wait until hands dry completely! 
Note
1. Please do not touch on the strip by your finger or use 
    the smearing blood sample.
2. Do not remove the strip before complete 6 seconds 
    count down, or it will be automatically shut down. 
3. The test strips cannot be re-used.
Blood glucose 
    120 mg/dL.
Now Testing.
Contr ol Solut ion M ode.
Please apply blood 
into the test strip.
1. Inserting the test strip.
3. Put your blood drop to the Reaction Area on the strip.
4. After dropping the blood , you will hear a beep sound, then it 
    starts  6 seconds count down at the same time.
5. The result will appear on the display, 
     auto-saving in memory.
6.Remove the used test strip, then it goes to sleep mode.
2. Press “M” button to highlight your status 
    as “control mode     ”, “before meals       ”,
    “after meals     ” or in general mode without 
    labeling.
a) Pressing once “M” button highlights the 
    value with       as control solution mode. 
    (Testing results in control solution mode 
    will not store in memory.)
b) Pressing once “M” button highlights the 
    value with      as before meals status.
c) Pressing once “M” button highlights the 
    value with     as after meals status.
d) Pressing one more “M” button removes the 
    highlight with general mode.
LCD Display
36.4   44.4
x
Insert test strip here
Battery cover
Label
For Model : GM20E