Справочник Пользователя для MOCACARE Corporation 22102158LD

Скачать
Страница из 51
49
Appendix
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments 
listed below, and should only be used in such environments:
Immunity 
test
IEC 60601 
test level
Electromagnetic environment – guidance
Conducted 
RF 
IEC 61000-4-6
Radiated RF 
IEC 61000-4-3
3 Vrms 150 
kHz to 80 
MHz
3 V/m
80 MHz to 
2.5 GHz
Portable and mobile RF communications 
equipment should be used no closer to 
any part of the device, including cables, 
than the recommended separation 
distance calculated from the equation 
applicable to the frequency of the 
transmitter.
Recommended separation distance
d = 1.2 80 √p MHz to 800 MHz
d = 2.3 800 √p MHz to 2.5 GHz 
where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in metres (m).
Field strengths from fixed RF transmitters, 
as determined by an electromagnetic 
site survey, a should be less than the 
compliance level in each frequency range. 
Not 
Applicable
 
10 V/m
80 MHz to 
2.7 GHz
b Interference may occur in the 
vicinity of equipment marked 
with the following symbol:
Compliance
level