Справочник Пользователя для MOCACARE Corporation 22102158LD
49
Appendix
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
Immunity
Immunity
test
IEC 60601
test level
Electromagnetic environment – guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
Radiated RF
IEC 61000-4-3
3 Vrms 150
kHz to 80
MHz
3 V/m
80 MHz to
2.5 GHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 80 √p MHz to 800 MHz
Recommended separation distance
d = 1.2 80 √p MHz to 800 MHz
d = 2.3 800 √p MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency range.
Not
Applicable
10 V/m
80 MHz to
2.7 GHz
b Interference may occur in the
vicinity of equipment marked
with the following symbol:
Compliance
level