Increasing Stimulation Strength
The sleep remote allows you to increase stimulation 
strength within a range set by your doctor. If you feel little or 
no stimulation, you can increase (
) the stimulation. 
Stimulation can be increased when therapy is on or off.
The stimulation strength gauge on the back of the sleep 
remote indicates the current stimulation strength and the 
upper limit light (
) indicates when you have reached the 
highest stimulation strength allowed by your doctor. (For 
information on the stimulation strength gauge and upper 
limit, refer to page 18.) 
Contact your doctor if you can’t achieve the desired 
stimulation strength.
To test the current stimulation strength:
1. Turn therapy off by pressing the Therapy Off button (
and positioning the sleep remote over your generator. 
2. Wait until you hear a tone or 10 seconds pass.
3. Turn therapy on by pressing the Therapy On button (
and positioning the sleep remote over your generator.
4. Wait until you hear a tone or 10 seconds pass.
5. You will receive one short stimulation immediately when 
therapy is turned on.