Справочник Пользователя для Chauvin Arnoux P01120308Clip-on ammeter adapter 52 mm P01120308

Модели
P01120308
Скачать
Страница из 40
13
n
C160 : 1000 A calibre (2000 A max. peak)
Intensity in A~
1...50 A
50...200 A
200...1000 A
1000...1200 A
Intrinsic error 
(1)
 1% + 1 mV
Dephasing
(2)
 3°
 2°
 1°
C160 : 100 A calibre (300 A max. peak)
Intensity in A~
0.1...5 A
5...20 A
20...100 A
100...120 A
Intrinsic error 
(1)
 2% + 5 mV
Dephasing
(2)
 15°
 10°
 5°
C160 : 10 A calibre (30 A max. peak)
Intensity in A~
0.1...0.5 A
0.5...2 A
2...10 A
10...12 A
Intrinsic error 
(1)
 3% + 10 mV
Dephasing
(2)
(2)
 15°
 15°
(1) As % of output signal
(2) Unspecified
n
C173 : 1000 A calibre
Intensity in A~
1...10 A
10...100 A
100...1000 A
1000...1200 A
Intrinsic error 
(1)
 1% + 0.2 mV
 0,5% + 0.2 mV
 0.2%
 0.2%
Dephasing
(2)
 2°
 1°
 1°
C173 : 100 A calibre
Intensity in A~
0.1...1 A
1...10 A
10...100 A
100...120 A
Intrinsic error 
(1)
 1% + 0.2 mV
 0.5% + 0.2 mV
 0.3%
 0.2%
Dephasing
(2)
 2°
 1°
 1°
C173 : 10 A calibre
Intensity in A~
0.01...0.1 A
0.1...1 A
1...10 A
10...12 A
Intrinsic error 
(1)
 1% + 0.2 mV
 0.5% + 0.2 mV
 0.5%
 0.5%
Dephasing
(2)
 5°
 2°
 2°
C173 : 1 A calibre
Intensity in A~
0.001...0.01 A
0.01...0.1 A
0.1...1 A
1...1.2 A
Intrinsic error 
(1)
 3% + 1 mV
 3% + 1 mV
 0.7% + 1 mV
 0.7% + 1 mV
Dephasing
(2)
(2)
 10°
 10°
(1) As % of output signal
(2) Unspecified
4.4
Conditions of use
The "C100" clamps must be used in the following conditions, in order to comply with user safety and
metrological performance requirements.