RIGOL 
DS1000Z Programming Guide 
2-131
 
:TRIGger:SPI:DLEVel 
Syntax 
:TRIGger:SPI:DLEVel <level> 
:TRIGger:SPI:DLEVel? 
Description 
Set or query the trigger level of the SDA channel in SPI trigger. The unit is the same as the 
current amplitude unit. 
 
Parameter 
Name 
Type 
Range 
Default 
<level> 
Real 
(± 5 × VerticalScale from the screen center)  -
OFFSet 
 
Explanation 
For VerticalScale, refer to 
. For OFFSet, refer 
to 
 
Return 
Format 
The query returns the trigger level in scientific notation.   
Example 
:TRIGger:SPI:DLEVel 0.16   /*Set the trigger level to 160mV*/ 
:TRIGger:SPI:DLEVel?   /*The query returns 1.600000e-01*/ 