  DS1000Z Programming Guide 
:CHANnel<n> Commands 
 
:CHANnel<n> commands are used to set or query the vertical system parameters, such as bandwidth limit, 
coupling, vertical scale and vertical offset. 
 
Command List: 
 
 
:CHANnel<n>:BWLimit 
Syntax 
:CHANnel<n>:BWLimit <type> 
:CHANnel<n>:BWLimit? 
Description 
Set or query the bandwidth limit parameter of the specified channel. 
Parameter 
Name 
Type 
Range 
Default 
<n> 
Discrete 
{1|2|3|4} 
——
 
<type> 
Discrete 
{20M|OFF} 
OFF 
 
Explanation 
  OFF: disable the bandwidth limit and the high frequency components of the signal 
under test can pass the channel. 
20M: enable the bandwidth limit and the high frequency components that exceed 20 
MHz are attenuated.   
  Enabling the bandwidth limit can reduce the noise, but also can attenuate the high 
frequency components. 
Return 
Format 
The query returns 20M or OFF. 
Example 