:MEASure:SOURce 
Syntax 
:MEASure:SOURce <sour> 
:MEASure:SOURce? 
Description 
Set or query the source for parameter measurement.   
Parameter 
Name 
Type 
Range 
Default 
<sour> 
Discrete 
{CHANnel1|CHANnel2|CHANnel3|CHANnel4} 
CHANnel1 
 
Explanation 
This command is used to set the source for parameters (except for delay and phase) 
measurement. Delay and phase measurements need to set two sources which can be 
respectively set by using 
an
.   
Return 
Format 
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.   
Example 
:MEASure:SOURce CHANnel2 
Related 
Command 
 
 
:MEASure:COUNter:SOURce 
Syntax 
:MEASure:COUNter:SOURce <sour> 
:MEASure:COUNter:SOURce? 
Description 
Set or query the source of the frequency counter, or disable the frequency counter.   
Parameter 
Name 
Type 
Range 
Default 
<sour> 
Discrete  {CHANnel1|CHANnel2|CHANnel3|CHANnel4|OFF}  OFF 
 
Return 
Format 
The query returns CHAN1, CHAN2, CHAN3, CHAN4 or OFF.   
Example 
:MEASure:COUNter:SOURce CHANnel2 
Related 
Command 
 
 
 
:MEASure:COUNter:VALue? 
Syntax 
:MEASure:COUNter:VALue? 
Description 
Query the measurement result of the frequency counter. The default unit is Hz.   
Return 
Format 
The query returns the measurement result in scientific notation. If the frequency counter 
is disabled, 0.0000000e+00 will be returned.   
Example 
:MEASure:COUNter:VALue?   /*The query returns 1.000004e+03*/ 
Related 
Command