Operator's Manual
920836 Rev A
43
DISPLAY - PERSIST_SETUP, PESU
Description
The PERSIST_SETUP command sets the duration of the display, in seconds, when in persistence
mode. The persistence can be set on all traces or per trace.
mode. The persistence can be set on all traces or per trace.
The PERSIST_SETUP? query indicates the current status of the persistence.
Command Syntax
PErsist_SetUp <time>
<time>:= {1, 2, 5, infinite}
NOTE: The <mode> argument Top2 used by some older instruments is not supported.
Query Syntax
PErsist_SetUp?
Response Format
PErsist_SetUp <time>
Example
The following instruction sets the variable persistence to 10 seconds on all traces:
PESU 20
Related Commands
PERSIST