Справочник Пользователя для Rigol DS1074Z 4-channel oscilloscope, Digital Storage oscilloscope, DS1074Z
Модели
DS1074Z
RIGOL
DS1000Z User’s Guide
13-3
Storage Type
Press Storage Storage to select the desired storage type. The default is
“Picture”. The storage and recall descriptions of each type are as follows.
1. Picture
Save the screen image in external memory in “.PNG”, “.BMP8”, “.BMP24”,
“.JPEG” or “.TIFF” format. You can specify the file name and saving directory
and save the corresponding parameter file (.txt) under the same directory using
the same file name and the function of this parameter file is the same to the
“.JPEG” or “.TIFF” format. You can specify the file name and saving directory
and save the corresponding parameter file (.txt) under the same directory using
the same file name and the function of this parameter file is the same to the
. The recall of image and parameter files is not supported.
After seleting this type:
Press Pic Type to select the desired storage format.
Press Para.Save to enable or disable the parameter save function.
Press Pic Type to select the desired storage format.
Press Para.Save to enable or disable the parameter save function.
One-key Bitmap Saving
After a USB storage device is connected, press
at the front panel to
quickly save the current screen image under the root directory of the USB
storage device in “png” format.
2. Traces
Save the waveform data in external memory in “.trc” format. The data of all the
channels turned on can be saved in the same file. At recall, the data will be
displayed on the screen directly.
channels turned on can be saved in the same file. At recall, the data will be
displayed on the screen directly.
3. Waveforms
Save the waveform data in external memory in “.wfm” format. The stored files
contain the waveform data of the two analog channels and the main setting
information of the oscilloscope and all the data can be recalled.
contain the waveform data of the two analog channels and the main setting
information of the oscilloscope and all the data can be recalled.
4. Setups
Save the settings of the oscilloscope in internal or external memory in “.stp”
format. The stored settings can be recalled.
format. The stored settings can be recalled.
5. CSV
Save the waveform data displayed on the screen or of the specified channels in