Техническая Спецификация для Dostmann Electronic Dostmann electronic Temperature Data Logger 5005-1105
Модели
5005-1105
LOG100/110/CRYO
– Data logger
G
16
H: Measurement 2 displays the measurement in the lower display line.
Depending on the logger model, the settings of the internal or the
external temperature measurement, average, minimum or maximum
measurements will be displayed.
external temperature measurement, average, minimum or maximum
measurements will be displayed.
I:
Unit Measurement 2 display the current measuring unit of measurement
2.
2.
J:
MAX-MIN-AVG display the average, minimum or maximum measure-
ments.
ments.
K: Status info display the operation mode LOG or STOP. LOG indicates the
recording mode and STOP indicates standby mode.
L:
External probe EXT s displayed when an external sensor is connected. In
the lower display line the measurement 2 is corresponding with the
external sensor.
the lower display line the measurement 2 is corresponding with the
external sensor.
M: Lowbat indicates the capacity of the battery.
Note:
°C = Celsius,°F = Fahrenheit
%rh = relative humidity, td = dew point temperature
%rh = relative humidity, td = dew point temperature
Other display information
• In addition to the above mentioned information, the display also indicates
• In addition to the above mentioned information, the display also indicates
several other information. This information will be displayed depending
on the display settings (snooze function) and operation mode:
on the display settings (snooze function) and operation mode:
Display
connected
Logger
switched off
to the PC
configured
LOG100/110/CRYO
– Data logger
G
13
1.3 Standard settings / Factory settings
• Note the following default settings of the data logger before first use. By
• Note the following default settings of the data logger before first use. By
using the DE-LOG-Graph software, the setting parameter can easily be
changed:
1.
changed:
1.
Description: blank (max. 16 characters)
2.
LCD-Snooze mode: µ
3.
LCD-Snooze after Sec.: 10
4.
Mode-button active: µ
5.
Alarm settings for temperature: µ
-30.0°C 70.0°C
-40.0°C 150.0°C
Alarm settings humidity: µ
6.
Alarm delay: □
7.
Alarm cumulation: off
8.
Alarm reset: µ
9. Alarm-LED-interval 3 seconds
10. Alarm-LED-blink duration 1 second
11. Alarm-Buzzer-interval 30 seconds
12. Alarm-Buzzer-duration 1 second
13. Temperature unit: °C
14. Start-button active: µ
15. Start by Reed-contact: □ (by request only)
16. Waiting for manual start: µ
17. Single use only: µ
18. Measuring interval: 15 Minutes
19. Stop-button active: µ
20. Stop by Reed-contact: □ (by request only)
21. Cycle memory: µ (if the memory is full the oldest measurement will be
10. Alarm-LED-blink duration 1 second
11. Alarm-Buzzer-interval 30 seconds
12. Alarm-Buzzer-duration 1 second
13. Temperature unit: °C
14. Start-button active: µ
15. Start by Reed-contact: □ (by request only)
16. Waiting for manual start: µ
17. Single use only: µ
18. Measuring interval: 15 Minutes
19. Stop-button active: µ
20. Stop by Reed-contact: □ (by request only)
21. Cycle memory: µ (if the memory is full the oldest measurement will be
overwritten)
µ
= Default
LOG100/110/CRYO
13
page
121023
Manual
0.0%
100.0%
Cryo:
350.0°C
70.0°C
-200.0°C
-10.0°C