Справочник Пользователя для Fluke Fluke 4325181

Модели
4325181
Скачать
Страница из 56
1625-2 
Users Manual 
12 
Table 5. Display Descriptions (cont.) 
Function Displays 
Condition 
Note 
 
 
edw042.eps 
Compensation higher 
than measured value. 
Delete compensation 
or switch instrument 
ON/OFF. 
 
 
edw043.eps 
Wrong polarity on 
jacks E and ES. 
Reverse polarity. 
 
 
edw044.eps 
Measured value 
unstable. 
Unsteady noise 
voltage. Try time 
average measurement. 
 
 
edw045.eps 
Current in external 
transformer too low. 
Reduce auxiliary 
current spike 
resistance. 
 
No reactions to button 
control etc.
 
edw046.eps 
Operation under faulty 
conditions. 
Check batteries. 
Switch ON/OFF if still 
faulty, contact service.
 
After 
"START" 
 
edw047.eps 
Reverse orientation of 
current clamp or 
"upwards" current. 
Reverse clamp. 
 
 
edw048.eps 
Checksum of EE 
PROM incorrect.